SRFD 2500Technical Parameters
Designation |
Technical Data |
Range of scanning (mm) |
Range of scanning (mm):0~6000
Steps: 2.5,5,10,20, 30,40,50,60,70,80,90, 100,
150,200, 250, 300, 350, 400, 450,500,600,700,
800,900,1000,2000,3000,4000,5000,6000
Adjusting step: 0.1mm(2.5 mm~99.9mm),
1mm(100mm~6000mm) |
D-delay (ms) |
D-delay (ms):-20~+3400
Steps: -20,-10,0.0, 10, 20, 50,100,150,200,250,
300,350,400,450,500, 600,700,800,900,1000,1500,
2000,2500,3000,3400.
Adjusting steps: 0.1(-20ms~999.9ms),1(1000ms~3400ms) |
P-delay (ms) |
P-delay:0.0~99.99
Adjusting steps: 0.01 |
MTLVEL(m/s) |
MTLVEL:1000~15000
7 fixed levels: 2260,2730,3080,3230,4700,5920,6300,12000
Adjusting steps: 1 |
Working mode |
Single probe (receiving and sending),
double probe (one for receiving and
another for sending), transmission (transmission probe) |
Frequency Range (MHz) |
0.5~20 |
Gain adjustment (dB) |
0~110
Adjusting step: 0.0, 0.1, 0.5, 1, 2, 6, 12 |
Reject |
0%~80% of screen height, step: 1% |
Vertical linear error |
Vertical linear error is not more than 3% |
Horizontal linear error |
Not more than 0.2% in the scanning range |
Sensitivity Leavings |
>/=62 dB |
Dynamic range |
>/=34dB |
Alarm |
Three modes, i.e. forbidden wave, loss wave and auto |
A-Scan display area |
Full screen or local
A-Scan display freezing and de-freezing A-Scan filling |
Data save |
300 A-Scan images (including setting of instrument) |
Standard communication
interface with PC |
RS232/USB |
Measuring unit |
mm/inch |
Battery |
Li battery 7.4V 4800mAh |
Power adaptor |
Input 100V~240V/50Hz~60Hz
Output 9VDC/1.5A |
Working temperature |
-10°C~50°C |
Working humidity |
20%~90% |
Port type |
C9 |
Overall dimension (mm) |
240*150*50 |
Weight (kg) |
1.6 |
Options
Table 1.2 List of options
Designation |
Quantity |
Serial cable |
1 (9 pin) |
Communication software for PC |
1 set |
|